Class II — GE Healthcare, LLC

Recalling firmGE HealthCare Technologies Inc.
Registered facilityView this company's 50 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonGE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.
ProductVivid E95/E90/E80, H45581DC, H45581LB and H45581DA.
Quantity971(412 US; 559 OUS)
DistributionUnited Kingdom, United Arab Emirates, Turkey, Thailand, Switzerland, Sweden, Spain, South Africa, Slovenia, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Norway, New Zealand, New Caledonia, Netherlands, Myanmar, Kuwait, Korea (Republic of), Japan, Italy, ISRAEL, Ireland, Iran, India, Hungary, Hong Kong, Guatemala, Greece, Great Britain, Germany, France, Finland, Estonia, Eg…
Initiated2017-03-10
Classified2017-04-11
Terminated2018-03-12
Recall numberZ-1779-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — GE Healthcare, LLC (Z-1779-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76805/

Not open yet

Alerts for GE HealthCare Technologies Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.