Class II — Acumed LLC
| Recalling firm | ACUMED LLC |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed. |
| Product | 6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixatio… |
| Quantity | 15 units |
| Distribution | Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain |
| Initiated | 2017-01-19 |
| Classified | 2017-02-14 |
| Terminated | 2017-05-08 |
| Recall number | Z-1199-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Acumed LLC (Z-1199-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76332/
Alerts for ACUMED LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.