RegWardenFDA facility compliance records
Recall

Class II — Fresenius Medical Care Renal Therapies Group, LLC

Recalling firmFresenius Medical Care Renal Therapies Group, LLC
Registered facilityView this company's 11 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhen the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
ProductFresenius 2008K2 Series : Hemodialysis System
Quantity60,151 units (US) 1773 (OUS)
DistributionNationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Initiated2016-12-21
Classified2017-01-13
Terminated2023-03-17
Recall numberZ-1028-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Fresenius Medical Care Renal Therapies Group, LLC (Z-1028-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76037/

Not open yet

Alerts for Fresenius Medical Care Renal Therapies Group, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.