Class II — Medimaps Group

Recalling firmMedimaps SA
Registered facilityMedimaps Group SA (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe FRAX adjusted for TBS values are not correct when: The FRAX feature is activated in TBS iNsight; and TBS has been computed from a spine scan where some vertebrae were excluded.
ProductTBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a medical device software that is installed on bone densitometers for analysis of bone microarchitecture and osteoporosis management.
Quantity15 units in US
DistributionWorldwide Distribution - US Nationwide - in the states of: IL, NC, CA, DE, NY, MD, WI, MN. And the following countries: France,Thailand, Croatia, Canada, Italy, Spain, Russian Federation, Ukraine, Lithuania, South Africa, Brazil, Lebanon, Saudi Arabia, Korea, China Portugalm Iraq, Syrian Arab Republic, Czech Republic, Australia, Belgium, Austria, Romania, New Zealand, Switzerland, Hong Kong, Swede…
Initiated2016-10-03
Classified2016-11-05
Terminated2017-06-07
Recall numberZ-0369-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Medimaps Group (Z-0369-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75510/

Not open yet

Alerts for Medimaps SA

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.