RegWardenFDA facility compliance records
Recall

Class II — Mevion Medical Systems, Inc.

Recalling firmMevion Medical Systems, Inc.
Registered facilityMEVION MEDICAL SYSTEMS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonSoftware defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the radiograph with an error that increases with the distance of the structure from isocenter
ProductMEVION S250 Product Usage: Proton Radiation Therapy System
Quantity6 units
DistributionUS Nationwide Distribution in the states of: FL, MO, NJ, OH, OK
Initiated2016-10-20
Classified2016-11-14
Terminated2017-05-15
Recall numberZ-0411-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Mevion Medical Systems, Inc. (Z-0411-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75494/

Not open yet

Alerts for Mevion Medical Systems, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.