Class II — Hand Biomechanics Lab Inc

Recalling firmHAND BIOMECHANICS LAB, INC.
Registered facilityHAND BIOMECHANICS LAB, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA number of Agee WristJack® Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.
ProductAgee-WristJack® Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. The non-perishable sterile system is packaged in an inner then outer tray and shipped in a cardboard shelf box. Orthopedic:The WristJack Fracture Reduction System is a fracture reduction and exte…
Quantity65 units
DistributionNationwide Distribution
Initiated2016-09-14
Classified2016-10-18
Terminated2016-11-15
Recall numberZ-0136-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Hand Biomechanics Lab Inc (Z-0136-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75255/