Class II — Argon Medical Devices, Inc

Recalling firmARGON MEDICAL DEVICES, INC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheath
ProductMicroStic Plus w/SS Wire 5F and MicroStic Plus w/NiTi Wire & Echo Needle 4F. A sheath used to facilitate placing a catheter through the skin into a vein or artery.
Quantity200 units
DistributionWorldwide Distribution -- US, to the states of TN and MS; and the country of Canada.
Initiated2016-09-08
Classified2016-10-06
Terminated2017-06-07
Recall numberZ-0051-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Argon Medical Devices, Inc (Z-0051-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75213/

Not open yet

Alerts for ARGON MEDICAL DEVICES, INC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.