Class II — Roche Diagnostics Operations, Inc.

Recalling firmRoche Diagnostics Operations Inc.
Registered facilityView this company's 14 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonRoche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Productcobas e411 Immunoassay Analyzer e411: Immunoassay Analyzer Product Usage: The cobas p501/p701 instrument stores primary and secondary sample tubes. The tubes are stored in a controlled, refrigerated sample store. Automatic retrieval for add-on testing is possible at any time. Expired samples are disposed of automatically when their storage time has elapsed. The instrument is intended as accessory for diagnostic use.
Quantity9014 in total
DistributionUS Nationwide Distribution
Initiated2016-07-15
Classified2016-10-25
Terminated2017-06-05
Recall numberZ-0095-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Roche Diagnostics Operations, Inc. (Z-0095-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75186/

Not open yet

Alerts for Roche Diagnostics Operations Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.