Class II — Spiegelberg Gmbh & Co. KG

Recalling firmSpiegelberg GmbH & Co. KG
Registered facilitySPIEGELBERG GmbH & Co. KG (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonIn the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.
ProductSpiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.
Quantity22 units
DistributionNationwide Distribution to PA only
Initiated2009-12-29
Classified2016-09-26
Terminated2016-10-05
Recall numberZ-2888-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Spiegelberg Gmbh & Co. KG (Z-2888-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75160/

Not open yet

Alerts for Spiegelberg GmbH & Co. KG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.