Class II — R & D Systems, Inc.

Recalling firmR & D SYSTEMS, INC.
Registered facilityR & D SYSTEMS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonR&D Systems, Inc. received two customer complaints that results of the Controls provided with the Quantikine® IVD® Human sTfR Immunoassay kit were out of the range high. Internal testing within R&D Systems confirmed a problem with Control results being out of range high.
ProductR&D Systems, Quantikine® IVDsTfR Immunoassay Kit, for the quantitative determination of soluble transferrin receptor (sTfR) concentrations in human serum and plasma For in Vitro Diagnostic Use, REF DTFR1.
Quantity725 kits (502 US, 223 OUS)
DistributionUS: CA, CO, KY, LA, MD, MA, NJ, OH, RI, UT, VA, WA. OUS: AUSTRALIA, China, INDIA, INDONESIA, ISRAEL, JAPAN, MEXICO, SOUTH KOREA, UNITED KINGDOM.
Initiated2016-08-24
Classified2016-09-26
Terminated2017-07-26
Recall numberZ-2887-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — R & D Systems, Inc. (Z-2887-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75016/

Not open yet

Alerts for R & D SYSTEMS, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.