Class II — BBI SOLUTIONS OEM LTD

Recalling firmBBI Solutions OEM Limited
Registered facilityBBI SOLUTIONS OEM LIMITED (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.
ProductSimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood
Quantity420 units (183 units quarantined)
DistributionUS distribution to CT and IN
Initiated2016-06-01
Classified2016-10-27
Terminated2018-06-14
Recall numberZ-0314-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — BBI SOLUTIONS OEM LTD (Z-0314-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74786/

Not open yet

Alerts for BBI Solutions OEM Limited

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.