Class II — Intuitive Surgical, Inc.
| Recalling firm | Intuitive Surgical, Inc. |
|---|---|
| Registered facility | View this company's 41 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend |
| Product | STAPLER 45, IS4000, model number 470298 Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist® Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line or tissue butressing material (natural or synthetic). |
| Quantity | 2065 total |
| Distribution | US, including Puerto Rico and Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Monaco, Netherlands, Norway, Portugal, , Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Venezuela. |
| Initiated | 2016-07-14 |
| Classified | 2016-08-05 |
| Terminated | 2017-01-03 |
| Recall number | Z-2348-2016 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Intuitive Surgical, Inc. (Z-2348-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74705/
Alerts for Intuitive Surgical, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.