Class II — RAYSEARCH LABORATORIES AB
| Recalling firm | RaySearch Laboratories AB (publ) |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | An issue was found with the evaluation of biological clinical goals in RayStation 3.0, RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5. Biological clinical goals for an adapted plan based on another planning CT than the original planning CT will show incorrect values. -- To the best of the firm's knowledge, the issue has not caused any patient mistreatment or other incidents. However, the user must be aware of the following information to avoid incorrect plan evaluation during treatment planning. |
| Product | RayStation 3.0, RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5 --- Radiation Therapy Treatment Planning System --- designed for treatment planning and analysis of radiation therapy, provides treatment unit set up parameters and estimates dose distributions. |
| Quantity | US: 170 units |
| Distribution | California, Connecticut, Delaware, Florida, Hawaii, Maine, Missouri, New York, Ohio, Texas and Washington |
| Initiated | 2016-07-01 |
| Classified | 2016-07-15 |
| Terminated | 2017-09-25 |
| Recall number | Z-2206-2016 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — RAYSEARCH LABORATORIES AB (Z-2206-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74618/
Alerts for RaySearch Laboratories AB (publ)
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.