Class II — Siemens Healthcare Diagnostics Inc

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSoftware defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours.
ProductStratus® CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531
Quantity977 units
DistributionWorldwide Distribution -- USA, U.A.E., Algeria, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Canada, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Libya, Lithuania, Malaysia, Mexico, Namibia, P.R. China, Poland, Portugal, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan…
Initiated2016-06-22
Classified2016-08-11
Terminated2018-11-29
Recall numberZ-2525-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics Inc (Z-2525-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74534/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.