Class II — Esaote S.p.A.

Recalling firmESAOTE S.P.A.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe device is being recalled because dynamic loading forces from larger patients can fracture the extremity portion of the patient table. Additionally, a weight limit is added to the patient seat for weight-bearing examinations.
ProductMagnetic Resonance System G-scan Brio (Part Number /Model No.101002000)
Quantity11 USA; 29 OUS
DistributionDomestic: FL, TX, OH, MN, NV; Foreign: No customers in Canada; No VA/DOD.
Initiated2016-06-06
Classified2016-07-05
Terminated2017-05-30
Recall numberZ-2124-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Esaote S.p.A. (Z-2124-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74391/