RegWardenFDA facility compliance records
Recall

Class II — Ziehm Imaging Inc

Recalling firmZIEHM IMAGING GMBH
Registered facilityZIEHM IMAGING GMBH (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonMalfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
ProductZiehm Vision RFD 3D Mobile C-arm for Mobile Fluoroscopic Imaging
Quantity31 devices
DistributionUS Distribution including Puerto Rico and to the states of :FL, GA, VA,MA. ME, MI, NY, NJ, PA, RI, IL, TX, AZ, CA, NV, UT, and WA.
Initiated2016-04-15
Classified2016-08-25
Terminated2019-07-18
Recall numberZ-2639-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ziehm Imaging Inc (Z-2639-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74013/

Not open yet

Alerts for ZIEHM IMAGING GMBH

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.