Recall

Class II — Bradshaw Medical Inc.

Recalling firmBradshaw Medical Inc.
Registered facilityBradshaw Medical Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting.
ProductAxial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw Medical Axial Torque Limiting Driver is a reusable surgical instrument designed to limit the magnitude of applied (input) torque transferred to an external (output) device or fastener in the clockwise direction. The device is a non-adjustable mechanical click style hand torque tool which presents an audible and tactile click when the maximum desired torque is reached.
Quantity3,760 units
DistributionUS Nationwide Distribution in the states of: CA, FL, IN, MA, MI, OH, PA, TN UT, VA and the country of UNITED KINGDOM
Initiated2015-02-02
Classified2016-05-10
Terminated2020-01-10
Recall numberZ-1637-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Bradshaw Medical Inc. (Z-1637-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73873/

Not open yet

Alerts for Bradshaw Medical Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.