Recalling firmViewray Systems, Inc.
Registered facilityVIEWRAY SYSTEMS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonWhen editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a result there is the potential to deliver dose to the initial isocenter rather than the new location.
ProductMRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian® System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated
Quantity5 units
DistributionWorldwide Distribution - US including CA, FL, MO & WI., and International distribution to the Republic of Korea and The Netherlands.
Initiated2016-03-07
Classified2016-03-31
Terminated2017-02-22
Recall numberZ-1309-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Viewray Incorporated (Z-1309-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73550/

Not open yet

Alerts for Viewray Systems, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.