Class II — Collagen Matrix Inc
| Recalling firm | COLLAGEN MATRIX, INC. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the device, 3 of 13 products did not pass the recovery test, which requires that the compressed product fully expand within 1 minute of hydration time. |
| Product | NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 22 mm diameter x 12 m height) Manufacturer:- Collagen Matrix, Distributor:- ACE Surgical Supply Co, Inc, Intended use for dental surgery such as -augmentation or reconstructive treatment of alveoloar ridge -filling of infrabony periodontal defects -filling of defects after root resection, apicoectomy and cystectomy -filling of extraction sockets to enhance preservation of the alveolar ridge -elevation of maxillary sinus floor -filling of periodontal defects in conjun… |
| Quantity | 347 units |
| Distribution | US Distribution to state of: MA. |
| Initiated | 2013-05-13 |
| Classified | 2016-03-11 |
| Terminated | 2017-06-21 |
| Recall number | Z-1097-2016 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Collagen Matrix Inc (Z-1097-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73206/
Alerts for COLLAGEN MATRIX, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.