Class I — Maquet Cardiovascular Us Sales, Llc
| Recalling firm | Maquet Cardiovascular LLC |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Terminated |
| Reason | Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US. |
| Product | Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration. |
| Quantity | 7 units within US |
| Distribution | US Nationwide in the states of CA, NY, and OR |
| Initiated | 2015-12-10 |
| Classified | 2016-02-22 |
| Terminated | 2016-07-06 |
| Recall number | Z-0804-2016 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Maquet Cardiovascular Us Sales, Llc (Z-0804-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73137/
Alerts for Maquet Cardiovascular LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?