RegWardenFDA facility compliance records
Recall

Class I — Boston Scientific Corporation

Recalling firmBoston Scientific Corporation
Registered facilityView this company's 37 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonBoston Scientific is voluntarily recalling its Chariot" Guiding Sheath. To date, Boston Scientific has received fourteen complaints for shaft separation, four for distal shaft separation. The most severe outcome of this failure is embolism of device fragments.
ProductBoston Scientific, Chariot ST-CCV Guiding Sheath, Sterilized using ethylene oxide: Product Description Material Number (UPN) Guiding Sheath 5F, 45 cm, ST, CC H74939277545110 Guiding Sheath 5F, 45 cm, MP, CC H74939277545210 Guiding Sheath 6F, 45 cm, ST, CC H74939277645110 Guiding Sheath 6F, 45 cm, ST, TB H74939277645120 Guiding Sheath 6F, 45 cm, MP, CC H74939277645210 Guiding Sheath 6F, 45 cm, MP, TB H74939277645220 Guiding Sheath 6F, 65 cm, ST, CC H74939277665110 Guiding Sheath 6F, 65 cm, ST, TB H74939277665120 Guiding Sheath 6F, 90 cm, ST, CC H74939277690110 Guiding Sheath 6F, 90…
Quantity7985
DistributionWorldwide Distribution - US (nationwide) including DC and Puerto Rico, and in the states of AK, AL, AR. AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY and the countries of Austria, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Hong Kong, Hunga…
Initiated2015-11-19
Classified2015-12-18
Terminated2017-02-23
Recall numberZ-0392-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Boston Scientific Corporation (Z-0392-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/72702/

Not open yet

Alerts for Boston Scientific Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.