RegWardenFDA facility compliance records
Recall

Class II — DePuy Spine, Inc.

Recalling firmDePuy Spine, Inc.
Registered facilityDEPUY SPINE, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonManufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length
ProductAegis Torque Handle CATALOG NO. : 2871-10-800. A spinal surgical instrument used on the AEGIS Anterior Lumbar Plate System.
Quantity16 devices
DistributionUS Distribution to the states of : NC, MI, CO, MD, FL, ID, TX, CA and NY.
Initiated2015-11-20
Classified2015-12-21
Terminated2017-07-03
Recall numberZ-0454-2016
InitiationVoluntary: Firm initiated