RegWardenFDA facility compliance records
Recall

Class I — Event Medical LTD

Recalling firmeVent Medical
Registered facilityEVENT MEDICAL, LTD. (the firm's only registered facility)
ClassificationClass I
StatusTerminated
ReasonA component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.
ProductInspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport.
Quantity6441 units total (251 units in US)
DistributionWorldwide Distribution - US (Nationwide) and Internationally to: Algeria Argentina Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bolivia Canada Chile China Colombia Czech Rep Ecuador Egypt England Estonia Finland Germany Ghana Gibraltar Greece Guatemala Holland Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya …
Initiated2015-10-13
Classified2015-11-19
Terminated2017-06-22
Recall numberZ-0230-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Event Medical LTD (Z-0230-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/72491/

Not open yet

Alerts for eVent Medical

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.