RegWardenFDA facility compliance records
Recall

Class II — Viewray Incorporated

Recalling firmViewray Systems, Inc.
Registered facilityVIEWRAY SYSTEMS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly.
ProductMRIdian® ViewRay® Radiation Therapy System, ViewRay® Is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.
Quantity4 units
DistributionUS Distribution to the states of : CA, MO and WI., and Internationally to Korea.
Initiated2015-08-27
Classified2015-10-09
Terminated2017-06-09
Recall numberZ-0115-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Viewray Incorporated (Z-0115-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/72166/

Not open yet

Alerts for Viewray Systems, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.