RegWardenFDA facility compliance records
Recall

Class II — Zimmer Gmbh

Recalling firmZIMMER, INC.
Registered facilityView this company's 10 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonIn about 5% of tested samples, the outer tray was found to be compromised. Risks: A damaged outer tray detected during surgery may cause a delay in surgery to obtain another implant. If a compromised packaging is not detected prior or during a surgery by the user, it may lead to periprosthetic infection. This may result in revision or multi-stage revision to treat the infection.
ProductZimmer Natural Nail CM Long. Intramedullary Fixation Rod. The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone.
Quantity57,603 units
DistributionUS Nationwide distribution.
Initiated2015-09-03
Classified2015-10-14
Terminated2016-03-15
Recall numberZ-0119-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zimmer Gmbh (Z-0119-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/72123/