RegWardenFDA facility compliance records
Recall

Class II — Transonic Systems Inc

Recalling firmTRANSONIC SYSTEMS, INC.
Registered facilityTRANSONIC SYSTEMS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonSoftware defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
ProductTransonic Flow Probe, 3 mm, Catalog Number HQC3FMC -- Product Usage: to measure flow intra-operatively.
QuantityDomestic: 36 units
DistributionWorldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
Initiated2015-07-27
Classified2015-09-02
Terminated2016-07-18
Recall numberZ-2721-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Transonic Systems Inc (Z-2721-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71907/