RegWardenFDA facility compliance records
Recall

Class II — Roche Diagnostics Operations, Inc.

Recalling firmRoche Diagnostics Operations Inc.
Registered facilityView this company's 14 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonLow PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg
Productcobas b 123 Fluid Pack COOX REF 05170036001 400 Product Usage: The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes NAT, K+, iCaWt (ISE), hematocrit (THct), metabolites (GLu, Lac) total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COHb, MetHb), and oxygen saturation (SO2). The cobas b 123 POC system calculates derived parameters. Depending on the equipment configuration of the instrument, the Sensor Cartridge and the Fluid Pack used, the following parameters are measured in human whole blood and QC ma…
Quantity82 Fluid Packs
DistributionUS Nationwide Distribution in the states of LA, SC, CA, TX, NE, VA and WV including: Puerto Rico.
Initiated2015-07-14
Classified2015-08-21
Terminated2016-01-19
Recall numberZ-2447-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Roche Diagnostics Operations, Inc. (Z-2447-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71786/

Not open yet

Alerts for Roche Diagnostics Operations Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.