Class II — Pega Medical Inc.
| Recalling firm | Pega Medical Inc. |
|---|---|
| Registered facility | ORTHOPEDIATRICS CANADA ULC (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation. |
| Product | Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273. |
| Quantity | 4 units |
| Distribution | US Distribution to the states of : FL and TN. |
| Initiated | 2015-06-15 |
| Classified | 2015-06-26 |
| Terminated | 2015-12-29 |
| Recall number | Z-1848-2015 |
| Initiation | Voluntary: Firm initiated |