RegWardenFDA facility compliance records
Recall

Class II — ResMed Ltd.

Recalling firmRESMED CORP.
Registered facilityView this company's 7 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe Astral device allows clinicians to disable all alarms including those that detect circuit disconnection. Inappropriate disabling of alarms in ventilator dependent patients can lead to hazards of insufficient ventilation and insufficient expiratory pressure which can result in major severity harm.
ProductAstral 100 ( Model No. 27001) & Astral 150 (Model No. 27003) ventilator.
Quantity1519 units
DistributionWorldwide distribution. US nationwide, Uruguay and Canada.
Initiated2015-05-05
Classified2015-06-10
Terminated2016-04-07
Recall numberZ-1750-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ResMed Ltd. (Z-1750-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71100/