Class II — Pega Medical Inc.

Recalling firmPega Medical Inc.
Registered facilityORTHOPEDIATRICS CANADA ULC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.
ProductFemale Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval IM Telescopic System. For orthopedic procedures. This instrument is included in the Fassier-Duval IM Telescopic System instrument tray. This instrument is marked with the catalogue number and lot number.
Quantity9 units (5 units in the US and 4 units international)
DistributionWorldwide distribution. US in Delaware, Florida, Nebraska, and New York; Australia, Brazil, and United Kingdom.
Initiated2015-04-10
Classified2015-04-29
Terminated2015-06-29
Recall numberZ-1546-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Pega Medical Inc. (Z-1546-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71061/

Not open yet

Alerts for Pega Medical Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.