Class II — Viewray Incorporated
| Recalling firm | Viewray Systems, Inc. |
|---|---|
| Registered facility | VIEWRAY SYSTEMS, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The software was not correctly using the RT (Radiation Therapy) to MR (Magnetic Resonance image) coordinate correction for non HFS (Head First Supine) patient orientations, resulting in slice mismatch error. |
| Product | ViewRay System, Radiation Therapy System |
| Quantity | 1 |
| Distribution | US distribution to MO. |
| Initiated | 2015-01-15 |
| Classified | 2015-05-05 |
| Terminated | 2015-09-23 |
| Recall number | Z-1580-2015 |
| Initiation | Voluntary: Firm initiated |