Recall

Class II — Mindray DS USA, Inc. dba Mindray North America

Recalling firmMindray DS USA, Inc.
Registered facilityMINDRAY DS USA, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonSoftware anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
ProductPanorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.
Quantity4 units
DistributionWorldwide Distribution - US Nationwide in the state of TX.
Initiated2015-03-12
Classified2015-05-26
Terminated2017-03-28
Recall numberZ-1675-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Mindray DS USA, Inc. dba Mindray North America (Z-1675-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/70959/

Not open yet

Alerts for Mindray DS USA, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.