Class II — GE Healthcare

Recalling firmGE HealthCare Technologies Inc.
Registered facilityView this company's 50 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonAn issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
ProductGE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx
Quantity6 units in US
DistributionUS Distribution to the states of: AZ, FL, GA, MO and WI.
Initiated2014-12-19
Classified2015-04-27
Terminated2015-04-29
Recall numberZ-1406-2015
InitiationVoluntary: Firm initiated