Class II — SenTec AG
| Recalling firm | Sentec, Inc. |
|---|---|
| Registered facility | SENTEC, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | SenTec is voluntarily replacing V-Sign Sensors 2 due to pre-mature deterioration of a part of the PCO2 measuring unit and thus disabling PCO2-monitoring. This means that the affected sensors reach their end-of-life earlier than normally expected requiring the sensor`s replacement. There is no increased risk for the patient or operator, because the aforementioned deterioration results in the sens |
| Product | Sentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterproof, SenTec AG, CH-4106, Therwil, Switzerland. The SenTec Digital Monitoring System (SDMS) is used for continuous and non-invasive monitoring of transcutaneous carbon dioxide tension (tcPCO2), oxygen saturation (SpO2), and pulse rate (PR). |
| Quantity | 129 US, 59 OUS |
| Distribution | US Distribution to the state of : MO |
| Initiated | 2015-03-25 |
| Classified | 2015-04-22 |
| Terminated | 2016-02-05 |
| Recall number | Z-1501-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — SenTec AG (Z-1501-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/70878/
Alerts for Sentec, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.