RegWardenFDA facility compliance records
Recall

Class II — Biomerieux Inc

Recalling firmBIOMERIEUX SA
Registered facilityView this company's 16 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe test kits have Out-of-Range-Low (OORL) Flucytosine QC results for Candida krusei ATCC 6258
ProductVitek 2 AST-YS07 REF 414 967, Fungal Susceptibility Card, 20 cards per carton, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.
Quantity43,535/20-card cartons
DistributionDistributed only to foreign subsidiaries and distributors, including Canada and Mexico.
Initiated2015-03-27
Classified2015-05-12
Terminated2017-03-24
Recall numberZ-1602-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Biomerieux Inc (Z-1602-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/70868/