Class II — Roche Diagnostics Operations, Inc.
| Recalling firm | Roche Diagnostics Operations Inc. |
|---|---|
| Registered facility | View this company's 14 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Chemical component of the F3 Fluid pack and environmental bacterial contamination causes a positive bias up to 39% at 4.1 mmol/l for the glucose measurement and 30% at 4.12 mmol/l for lactate. Urea readings are unaffected. |
| Product | Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer |
| Quantity | 103 |
| Distribution | US Nationwide distribution. |
| Initiated | 2014-12-18 |
| Classified | 2015-02-02 |
| Terminated | 2017-02-06 |
| Recall number | Z-1041-2015 |
| Initiation | Voluntary: Firm initiated |