Class II — Ion Beam Applications S.A.
| Recalling firm | ION BEAM APPLICATIONS S.A. |
|---|---|
| Registered facility | ION BEAM APPLICATIONS S.A. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du |
| Product | adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device. |
| Quantity | 1 |
| Distribution | US Distribution in PA only. |
| Initiated | 2014-11-07 |
| Classified | 2014-12-19 |
| Terminated | 2015-06-29 |
| Recall number | Z-0836-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Ion Beam Applications S.A. (Z-0836-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69781/
Alerts for ION BEAM APPLICATIONS S.A.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.