Class II — Medtronic MiniMed Inc.
| Recalling firm | Medtronic |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Tubing may detach at the connect/disconnect location on the MiniMed Silhouette and MiniMed Sure-T infusion sets. Detachment causes interrupted insulin delivery and the pump will not alarm to notify your patient. This can lead to hyperglycemia, and diabetic ketoacidosis (DKA). |
| Product | Medtronic MiniMed Silhouette and/or MiniMed Sure-T Infusion Sets; Paradigm Silhouette infusion sets: MMT-368, MMT-368600, MMT-371, MMT-373, MMT-377, MMT-377600, MMT378, MMT-378600, MMT-38 1, MMT-381 600, MMT-382, MMT-382600, MMT-383, MMT-383600, MMT384, & MMT384600. Sure-T infusion sets MMT-862, MMT-863, MMT-864, MMT-865, MMT-866, MMT-873, MMT-874, MMTMMT-875, MMT-876, MMT-884, & MMT-886. MiniMed Duo sensor/infusion sets: MMT-5001A23, MMT-5001A43, MMT-5001B23, MMT-5001B43, MMT-5011A23, MMT-5011A43, MMT-5011B23, & MMT-5011B43. Silhouette and Sure-T disposable infusion sets are used with Medtro… |
| Quantity | US 12,400,098 and Internationally 8,266,733 |
| Distribution | Worldwide distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel,… |
| Initiated | 2014-11-04 |
| Classified | 2014-12-08 |
| Terminated | 2015-03-10 |
| Recall number | Z-0524-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Medtronic MiniMed Inc. (Z-0524-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69720/
Alerts for Medtronic
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?