Class III — Illumina Inc
| Recalling firm | Illumina, Inc. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class III |
| Status | Terminated |
| Reason | Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument. |
| Product | Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform. |
| Quantity | 19 units |
| Distribution | US: MD, NY Non-US: France |
| Initiated | 2014-11-06 |
| Classified | 2014-12-23 |
| Terminated | 2015-03-17 |
| Recall number | Z-0850-2015 |
| Initiation | Voluntary: Firm initiated |