Class II — Medical Positioning Inc
| Recalling firm | Medical Positioning, Inc. |
|---|---|
| Registered facility | MEDICAL POSITIONING, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction. |
| Product | Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to conduct syncope examinations. |
| Quantity | 17 beds |
| Distribution | Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI. |
| Initiated | 2014-07-10 |
| Classified | 2014-08-13 |
| Terminated | 2017-01-03 |
| Recall number | Z-2210-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Medical Positioning Inc (Z-2210-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68763/
Alerts for Medical Positioning, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.