RegWardenFDA facility compliance records
Recall

Class II — Boston Scientific Corporation

Recalling firmBoston Scientific Corporation
Registered facilityView this company's 37 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSome units of Intella Tip MiFi XP Temperature Ablation Catheters were not manufactured according to specification.
ProductIntellaTip MiFi XP Temperature Ablation Catheter; 8mm x 8F (2.67mm); Large Curve. The BSC Controller and Accessories are indicated for use in conjunction with standard and high power catheters for cardiac ablation procedures.
Quantity968 units total all models
DistributionNationwide Distribution.
Initiated2014-05-15
Classified2014-06-16
Terminated2015-06-25
Recall numberZ-1795-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Boston Scientific Corporation (Z-1795-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68373/

Not open yet

Alerts for Boston Scientific Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.