Class II — Integra LifeSciences Corp.
| Recalling firm | Integra LifeSciences Corporation |
|---|---|
| Registered facility | View this company's 15 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | This lot of Accell Evo3C Demineralized Bone Matrix, 5cc was incorrectly packaged in boxes that describe the product as Accell Evo3 Demineralized Bone Matrix. Evo3C is contraindicated for treatment of vertebral compression and not indicated for use in the posterolateral spine. |
| Product | Accell Evo3C Demineralized Bone Matrix, 5 cc, Model Number: 02-6000-050 , Lot Number: 132160. Product Usage: Accell Evo3c Demineralized Bone Matrix Putty is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone. |
| Quantity | 61 |
| Distribution | US Nationwide Distribution |
| Initiated | 2014-05-19 |
| Classified | 2014-06-24 |
| Terminated | 2014-09-05 |
| Recall number | Z-1868-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Integra LifeSciences Corp. (Z-1868-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68330/
Alerts for Integra LifeSciences Corporation
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.