RegWardenFDA facility compliance records
Recall

Class I — Convatec Inc.

Recalling firmConvatec
Registered facilityCONVATEC TRIAD LIFE SCIENCES, LLC (the firm's only registered facility)
ClassificationClass I
StatusTerminated
ReasonConvatec Inc. is recalling Flexi-seal FMS Control Kit since it does not have a 510K. In addition, an internal assessment of product performance, including a review of customer feedback, has confirmed that Flexi-Seal Control is not meeting expectations or those of customers. Specifically, the Auto-Valve feature that is unique to the Flexi-Seal Control has not been consistently performing relative
ProductConvatec Flexi Seal Control Fecal Management System Rx Only REF 411107 Product Usage:For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications as prescribed by a physician.
Quantity73,021 units US; 24,866 units OUS
DistributionWorldwide Distribution - USA Nationwide
Initiated2014-04-29
Classified2014-06-13
Terminated2015-08-28
Recall numberZ-1785-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Convatec Inc. (Z-1785-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68292/

Not open yet

Alerts for Convatec

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.