RegWardenFDA facility compliance records
Recall

Class II — LAP of America Laser Applications, LLC

Recalling firmLAP of America Laser Applications LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhile projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.
ProductLAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.
Quantity1 system
DistributionNorth Carolina
Initiated2014-03-21
Classified2014-06-24
Terminated2016-09-26
Recall numberZ-1866-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — LAP of America Laser Applications, LLC (Z-1866-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68259/

Not open yet

Alerts for LAP of America Laser Applications LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.