Class II — Heraeus Kulzer, LLC.

Recalling firmKulzer, LLC
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe firm discovered that the Gluma Desensitizer Power Gel can be inadvertently extruded from the application syringe all at once when it is supposed to extrude on the desired area a small amount at a time. If extruded all at once, product could reach the dental patients mucosa, face or eyes or the dental professional could be exposed. This would cause irritation, burning or blistering.
ProductGluma Desensitizer Power Gel Product Usage: Usage: This is a class 2 medical device. For reduction or elimination of pain in exposed cervical areas that do not require restoration. It is also used for alleviation or reduction of dentinal sensitivity after preparation of teeth to receive fixed prosthesis or restorations.
Quantity5,112 syringes or 1,278 kits (1 Kit=4 syringes)
DistributionUS Nationwide Distribution in the states of GA, TX, IL, CA, WA, WI, HI, MO, FL, NJ, TN, PA, NY, NC, OH, IN, CT, KY, AZ, NV, IA,
Initiated2014-04-23
Classified2014-05-22
Terminated2014-11-05
Recall numberZ-1651-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Heraeus Kulzer, LLC. (Z-1651-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68223/