Class II — Stryker Endoscopy

Recalling firmSTRYKER CORP.
Registered facilityView this company's 61 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonGravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
Product12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.
Quantity2591 trays
DistributionWorldwide distribution
Initiated2014-04-16
Classified2014-05-02
Terminated2014-08-18
Recall numberZ-1544-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Stryker Endoscopy (Z-1544-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/68045/

Not open yet

Alerts for STRYKER CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.