Class II — Philips Medical Systems (Cleveland) Inc
| Recalling firm | Philips Medical Systems International BV |
|---|---|
| Registered facility | View this company's 54 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | During internal testing, the Brilliance CT Big Bore was found to be out of tolerance for radio frequency emissions. Specifically, at the 48MHZ frequency, the testing indicated the Brilliance CT Big Bore was 3.5dB uV/meter higher than the applicable IEC 60601-1-2 standard specification. |
| Product | Brilliance CT Big Bore Oncology Whole Body Computed Tomography X-Ray System & Brilliance CT Big Bore Radiology Whole Body Computed Tomography X-Ray System. Product Usage: The Brilliance CT Big Bore is a Whole Body Computed Tomography X-Ray System which produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes. |
| Quantity | 454 Units |
| Distribution | Worldwide Distribution - US (nationwide) in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI & WV. and countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Ecuador, France, French Polynesia, Gabon, Germany… |
| Initiated | 2013-12-19 |
| Classified | 2014-06-21 |
| Terminated | 2015-08-25 |
| Recall number | Z-1745-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Philips Medical Systems (Cleveland) Inc (Z-1745-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67879/
Alerts for Philips Medical Systems International BV
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.