RegWardenFDA facility compliance records
Recall

Class II — Biocardia, Inc.

Recalling firmBioCardia
Registered facilityBIOCARDIA, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonFirm discovered the potential for particulates in the inner lumen of Morph Access Pro Steerable Introducer, which could result in am embolic event.
ProductMorph AccessPro Steerable Introducer; 6F Introducer, 90 cm working length: Single use only. BioCardia, Inc. San Carlos, CA 94070 BioCardia Morph AccessPro is intended to provide a pathway through medical instruments cush as balloon dilation catheters, guidewires or other therapeutic devices that may be introduced into the peripheral vasculature.
Quantity~2541 units - both catalog numbers.
DistributionWorldwide Distribution: US (nationwide) and country of: Israel.
Initiated2014-03-03
Classified2014-03-27
Terminated2014-06-19
Recall numberZ-1302-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Biocardia, Inc. (Z-1302-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67712/

Not open yet

Alerts for BioCardia

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.