RegWardenFDA facility compliance records
Recall

Class II — Lin-Zhi International Inc

Recalling firmLIN-ZHI INTERNATIONAL, INC.
Registered facilityLIN-ZHI INTERNATIONAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonProduct catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their shelf life due to use of dated but unexpired secondary stock used to create the products.
ProductCannabinoids DAU Calibrator Control Level 1 (62.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially infectious. IVD; The Cannabinoids (cTHC also known as THC version c) DAU negative and positive controls are human urine-based liquids and ready to use. The constituent is a processed drug-free human urine matrix, containing buffers, stabilizers, and less than 0.09 % of sodium azide. The controls are prepared by spiking known concentrations of 11-nor-9-THC-9-COOH into the drug-free matrix. cTHC is referenced as an abbreviation for 11-nor-9-THC-…
Quantity216 items
DistributionUS Distribution including the states of SC, NC, LA, FL, TX,, AL and GA and Internationally to Germany.
Initiated2014-02-06
Classified2014-03-12
Terminated2015-03-10
Recall numberZ-1179-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Lin-Zhi International Inc (Z-1179-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67521/