Class II — Zimmer, Inc.
| Recalling firm | ZIMMER, INC. |
|---|---|
| Registered facility | View this company's 10 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device. |
| Product | Natural-Knee® II - Unicompartmental Knee Prosthesis: CONG TIB INS SZ1/2 9MM UN CONG TIB INS SZ1/2 11MM U CONG TIB INS SZ1/2 13MM U CONG TIB INS SZ3/4 9MM UN CONG TIB INS SZ3/4 11MM U CONG TIB INS SZ3/4 13MM U CONG TIB INS SZ 5/6 9MM U CONG TIB INS SZ5/6 11MM U CONG TIB INS SZ5/6 13MM U |
| Quantity | 3638 units |
| Distribution | Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland. |
| Initiated | 2014-02-25 |
| Classified | 2014-03-20 |
| Terminated | 2015-03-24 |
| Recall number | Z-1251-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Zimmer, Inc. (Z-1251-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67503/
Alerts for ZIMMER, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.